>> Á¦Ç°¼Ò°³>> Àü±Ø&¾Ç¼¼»ç¸®
 
 
»ê¾÷¿ë Chloride (Cl-)
¼³¸í Industrial Chloride Ion Electrode, Combination, CPVC Body, 3/4"MPT
 
 
 

 

Model :

Slope :

# V-BC300-20H

52 ~ 59 mV / decade

Reproducibility :

+ / - 2%

Interferences :

S-2, I-1, CN-1, Br-1

Complexation :

Heavy metals

Temperature Range :

0 to 80¡Æ C (if calibrated at that temperature)

Pressure Range :

0 to 70 psi

ResponseTime :

95% response in 2 min

Storage :

Long term, store dry

Short term, store in dilute chloride standard

Concentration Range :

5 x 10-5 M to 1M/1.8 ppm to 35,500 ppm

pH Range :

2 to 12 pH

ISE Membrane Resistance :

< 25 Mohm

Reference Junction Resistance :
Reference Junction/Type :
Reference Inner / Outer Solution :
Care, Maintenance, & Cleaning :

< 50 Mohm
Ceramic Junction / Sealed
4M KCl/AgCl(1C)  /  1M KNO2(2IST-KNO3)
Polish with polishing strips Treat with a dilute ISA

Temperature Compensation :

Not recommended

Isopotential Point :

50,000 ppm Chloride

On-Line Capability :

Yes

Applications :

River and Tap Water, Plant Tissue, Soils, Boiler Feed Water, Blood Electrolytes/Clinical Analysis, Sweat, Urine, Blood, Cement, Plating Baths, Food Samples

 

 
¸ðµ¨¸í Ư¡ ±¸¸ÅÇϱâ